| Model | MT-101TH | MT-225TH | MT-504TH | MT-816TH | MT-1000TH | ||
|---|---|---|---|---|---|---|---|
| Workspace Volume(L) | 100 | 225 | 504 | 816 | 1000 | ||
| Workspace Dimensions (mm) | Width | 450 | 600 | 800 | 1000 | 1000 | |
| Deep | 450 | 500 | 700 | 800 | 1000 | ||
| Height | 500 | 750 | 900 | 1000 | 1000 | ||
| External Dimensions (mm) | Width | 1000 | 1150 | 1340 | 1550 | 1550 | |
| Deep | 1160 | 1210 | 1260 | 1370 | 1560 | ||
| Height | 1610 | 1870 | 2070 | 2170 | 2170 | ||
| Control model | Balanced Temperature&Humidity Control System(BTHC system) | ||||||
| Performance |
Temperature range (Optional) |
A: 0~150℃ B:-20~150℃ C:-40~150℃ D:-60~150℃ E:-70~150℃ | |||||
| Temperature fluctuation | ≤ ± 0.5 ℃ | ||||||
| Temperature uniformity | ≤ 2°C | ||||||
| Temperature deviation | ≤ ± 2°C | ||||||
| Heating rate | 1~3 °C/min | ||||||
| Cooling rate | 0.7~1 °C/min | ||||||
| Humidity range | 20~98% RH | ||||||
| Humidity deviation | ≤+2/-3%RH | ||||||
| Construction | Material | Exterior | A3 Steel board with powder coated | ||||
| Interior | Mirror Stainless Steel-SUS#304 | ||||||
| Thermal insulation | Glass fiber cotton or polyurethane foam | ||||||
| viewing window | Tempered glasses with conducting film | ||||||
| Air circulation | Fan type | Centrifugal Sirocco Fan | |||||
| Refrigeration | Cooling method | Single/Cascade refrigerating (air cooled) | |||||
| Compressor | Hermetically | ||||||
| Condenser | Air Cooled-plate fins with fan | ||||||
| Evaporator | Fin with several segments | ||||||
| Refrigerants | Environmentally friendly | ||||||
| Heating | Heater type | Nickel-chromium naked wire heating | |||||
| Humidifying system | Humidifying method | Shallow water bath with immersed stainless steel heater/sensor system to generate uniform humidity. | |||||
| Supplying water | Electromagnetic pump | ||||||
| tank capacity | 15L (Auto fill via pump) | ||||||
| Control and sense system | Controller | 7" color touch-screen | |||||
| Timing range | 1~9999 (H.M.S)adjustable | ||||||
| Temp. sensor | PT-100Ω Dry bulb | ||||||
| Humidity sensor | PT-100Ω Wet bulb | ||||||
| Facility requriement | Water | Deionized/distilled water for topping up water tank | |||||
| Electricity supply | 3 phase | ||||||
| Voltage (VAC) and frequency (Hz) dependant on country/region of installation. | |||||||
| Operating environment conditions | +5℃~ +28℃ ≤ 85%RH | ||||||
| Accessories | Cable access port soft plug; Specimen holdersWet bulb gauze; Fuses; Drain-pipe | ||||||
| Safety protection |
Over temperature protection; Compressor overload and high pressure protector; Over, absence and phase reverse voltage protector; Lack of water protector; Humidifier thermal overshoot protector Circuit breaker; Fire protection; Alarms; leakage protection; pressure relief/equalization vent |
||||||
Temperature test chambers focus on the required temperature simulation test from 4 optional low temperature -70 °C; -60°C; -40°C; -20°C to high temperature 180 °C.
Chambers with Single-stage Refrigeration systems are capable of achieving workspace temperatures down to -20°C. This is fine for many commercial and industrial tests that require low temperatures of typically 0°C, -10°C or -20°C. A single-stage system uses one compressor to directly cool the workspace.
Chambers with Cascade Refrigeration systems are capable of achieving workspace temperatures down to -40°C~-70°C depending on model. This is required for many ruggedized industrial, automotive and military test procedures that require low temperatures of typically -40°C, -55°C or -65°C. A cascade system uses two interdependent compressors.