Description
	Bench-top climatic chambers can fit on a personal desk, this small bench-top chamber achieves high performance to fulfill your climatic temperature humidity testing needs.
	One of the typical test conditions is known as 85/85, meaning 85°C and 85% of relative humidity, and the duration of the test may vary depending on used test method.
	
	Standard Features
	Cabinet 
	Interior stainless steel plate (sus#304)
	Baked panel and durable blue powder coated exterior steel plate
	High density polyurethane froth insulation or fiberglass insulation
	Welded structural base frame with removable service panels
	Dual silicone one-piece door gaskets with vulcanized corners
	Multilayer tempered heated viewing window; Anti-condensation & frost free
	One leak-proof 50mm (2 ") access port with plug on the left
	4 heavy duty casters
	Fin heat disperse-stainless steel heater
	PID controlled-stainless steel humidifier
	2 levels shelves, stainless steel
	Temperature measuring with PT100 in air outlet
	Humidity measuring with PT100 in air outlet
	Chamber drain
	Electrical 
	Anti-explosive interior lamps
	AC Power Cord
	Fast recovery fuses
	High pressure protect
	Phase reversal protect
	Circuit breaker system
	Lack of water detector
	Humidifier thermal overshoot protector
	System fault status indicators
	All wiring color coded and labeled for troubleshooting
	High temperature failsafe limit
	Humidity 
	Shallow water bath with immersed heater/sensor system to generate uniform humidity
	Vapor generator failsafe over temp protection
	Controls 
	Japanese Unique 7" color touch-screen controller
	Self prompting operation with 100 programmable segments.
	600 days’ historical date
	Real-time and historical curve record
	Ethernet Connection
	USB-Connection to save measuring data on USB Stick
	Online software to connect the PC with the controller to operate from PC
	Calibration 
	Calibration report for temperature humidity values from manufacturer
	Optional Accessories
	Additional cable access ports Ø 50 mm, 100 mm and 150 mm
	Additional shelves
	Glove ports to move the test specimen without opening the door
	Chart recorders
	Microprocessor (digital indicator) controller
	220, 415,440 or 480 Volts power input optional (depend on requirement)
	Humidity water purification system
	Additional Pt 100 for temperature measuring on test specimen
	Additional temperature protection for test specimen
	Smartphone or tablet pc WIFI monitoring
	Variable air speed in the test space
	Calibration certificate for temperature humidity values from 3rd measurement institution.
	RS-232/RS-485 data interface